TECHNICAL DATA

Test strip:

GlucoMen® areo Sensor

Assay Method:

Electrochemical, Glucose Oxidase

Icodextrin interference:

GlucoMen® areo system is suitable for use by people on peritoneal dialysis using solutions containing Icodextrin

Test Range:

20 – 600 mg/dL (1.1 – 33.3 mmol/L)

Haematocrit range:

10 – 70 %

Sample size:

Minimum 0.5 μL

Test time:

5 seconds

Batteries:

Two 3.0 V lithium batteries (CR2032)

Battery life:

A minimum of 1000 tests or about 1 year

Memory:

730 test results with markers, dates and times

Alarms:

Up to three settable acoustic alarms

Averages:

Over 1, 7, 14, 30, 60, 90 days

Dimensions:

85.5 mm (L) x 56 mm (W) x 18.2 mm (H)

Weight:

46 g (without batteries)

Meter Operating Conditions:

Temperature: 5 - 45 °C / (41 - 113 °F)
Relative Humidity: 20 - 90%